Needle deflection in thermal ablation procedures of liver tumors: a CT image analysis

Tonke L. de Jong, Camiel Klink, Adriaan Moelker, Jenny Dankelman, John van den Dobbelsteen

Research output: Chapter in Book/Conference proceedings/Edited volumeConference contributionScientificpeer-review

6 Citations (Scopus)
87 Downloads (Pure)

Abstract

Introduction: Accurate needle placement is crucial in image-guided needle interventions. A targeting error may be introduced due to undesired needle deflection upon insertion through tissue, caused by e.g. patient breathing, tissue heterogeneity, or asymmetric needle tip geometries. This paper aims to quantify needle deflection in thermal ablation procedures of liver tumors by means of a CT image analysis. Methods: Needle selection was done by using all clinical CT data that were made during thermal ablation procedures of the liver, ranging from 2008-2016, in the Erasmus MC, the Netherlands. The 3D needle shape was reconstructed for all selected insertions using manual segmentation. Subsequently, a straight line was computed between the entry point of the needle into the body and the needle tip. The maximal perpendicular distance between this straight line and the actual needle was used to calculate needle deflection. Results: In total, 365 needles were included in the analysis ranging from 14G to 17G in diameter. Average needle insertion depth was 95mm (range: 32 mm – 182 mm). Needle deflection was on average 1.3 mm (range: 0.0 mm – 6.5 mm). 54% of the needles (n=196) had a needle deflection of more than one millimeter, whereas 7% of the needles (n=25) showed a large needle deflection of more than three millimeters.
Conclusions: Needle deflection in interventional radiology occurs in more than half of the needle insertions. Therefore, deflection should be taken into account when performing procedures and when defining design requirements for novel needles. Further, needle insertion models need to be developed that account for needle deflection.
Original languageEnglish
Title of host publicationProceedings of SPIE
Subtitle of host publicationMedical Imaging 2018: Image-Guided Procedures, Robotic Interventions, and Modeling
EditorsB Fei, RJ Webster III
Place of PublicationBellingham, WA, USA
PublisherSPIE
Number of pages6
Volume10576
DOIs
Publication statusPublished - 2018
EventMedical Imaging 2018: Ultrasonic Imaging and Tomography - Houston, United States
Duration: 10 Feb 201815 Feb 2018

Publication series

NameProceedings of SPIE
Volume10576
ISSN (Electronic)1605-7422

Conference

ConferenceMedical Imaging 2018: Ultrasonic Imaging and Tomography
Country/TerritoryUnited States
CityHouston
Period10/02/1815/02/18

Keywords

  • bending
  • deflection
  • interventional radiology
  • MW
  • needles
  • RFA
  • thermal ablation

Fingerprint

Dive into the research topics of 'Needle deflection in thermal ablation procedures of liver tumors: a CT image analysis'. Together they form a unique fingerprint.

Cite this